permanence of a patient’s record, store the printed paper away from direct
sunlight, heat, or fumes from organic compound s. Storage temperatures
above 60°C (140°F), sustained exposu
re to direct sunlight, o r exposure to
high h umidity, acetone, ammonia, alcohols, or other organic compounds may
cause t he pap er to discolor.
NOTE: If printed reports are to be kept for prolonged periods, you must make
a photocopy of the thermosensitive pape r as this paper is not intended for
long-term retention and will los
e legibility over tim e.
NOTE: Contact with adhesive tape or page protectors will fade the printing
after 30 days.
Cleaning the PRM and Accesso ries
Clean the housing and touchscreen of the PR M with a soft clo th lightly
dampened with water, isopropyl alcohol, a 5% bleach solution, or window
cleaner.
Clean the ZWT ho using with a soft cloth lightly dampen ed with water, isopropyl
alcohol, a 5% bleach solution,
or window cleaner. Do not allow any amount of
cleaning solution o r moisture to come in contact with the USB port.
Clean t he printer/recorder wi
th a dry, soft brush to eliminate dust and particles
that may accumulate during printing and storage.
Clean the printer ro ller wit
h an alcohol wipe.
CAUTION: Do not use an abrasive cloth or volatile solvents to clean any
portion of the PRM or ZWT.
The cables used with the PRM are not sterile when packaged and cannot
be sterilized. When necessa
ry, clean the c ables with a soft cloth dampened
with a mild cleaning solution such as green soap, green soap tincture (U.S.
Pharmacopeia), Borax, or alcohol-free ha nd soap. Use a fresh soft cloth
dampened with sterile wat
er to remove residue. Towel-dry or air-dry the cables.
DO NOT use an ultraso nic cleaner. DO NOT immerse the cables. When
necessary, disinfect the ECG cable using a 2% glutaraldehyde solution (such
as Cidex) or a 10% bleach s
olution.
NOTE: Discard the ECG cables any time surface cracks appear in the
cables and/or the cables
discolor, become visibly worn, or if labeling becomes
unreadable.
Clean the sterilizabl
e telemetry wand in the same manner. DO NOT use an
ultrasonic cleaner. DO NOT immerse the telemetry wand. DO N OT allow
fluid to enter the wand cavity. Refer to "Preparing the PRM for Use" on page
6 fo r sterilization in
structions.
Patient Data Disk
ThePatientDataDiskcanbeusedtosavepatientdata. Becertainthatthe
write-protect tab is closed on the disk (Figure 10 on page 24). The write-protect
tabmustbeclosedin
order to record data to the d isk and to print reports. If
data cannot be re corded to the disk, check to see that the tab is positioned
to co ver the hole.