• Use the equipotential stud connection to equa lize the PRM’s galvan ic
potential with other electrical equipment. Using this terminal to
provide equalization m ay reduce elec
trical noise problems and the
possibility of indirect lea k age currents between the PRM and other
electrical equip ment.
CAUTION: Ensure the back o f the unit is accessible at all times so that
the power cord can b e disconnected.
• Plug the power cord into the appropriate AC outlet.
WARNING: To avoid the risk of ele c t
ric shock, only connect the PRM to
a grounded/earthed power source.
WARNING: Do not use the PRM or ZWT ad
jacent to or stacked with
other equipment. If adjacent or stacked use is necessary, check the PRM
and ZWT for normal operation in that configuration.
4. Start the PRM.
• Raise the screen to a comfortabl
e vi ewing angle.
• Press the On/Off button.
• Wait for the startup screen to app ear.
NOTE: During PRM startup, obser
ve the screen for any messages. If an
error message appears, write a detai led description of the error and contact
Boston Scientific using the information on the back cover.
Prepare for ZIP Telemetry
NOTE: The ZIP telemetry feature
is not available for all pulse generators.
For more information, refer to the associated product literature for the pulse
generator being interrogated.
1. For pulse generators that use the 3140 Zoom Wireless Transmitter to
communicate using ZIP telemetry:
a. For optimum ZIP telemetry communication, position the ZWT so that
it is no further than 3 m (10 ft) from the pulse generator and no closer
than7.6cm(3in)fromthePRM
.
b. Remove obstructions between the ZWT and the pulse generator.
NOTE: Repositioning the ZWT further away from the PRM may improve ZIP
telemetry performance. If ZIP telemetry pe rformance is not satisfactory, use
the teleme t ry wand instead
.
2. For pulse generators that do not use the 3140 Zoom Wireless Transmitter
to communicate using ZIP te
lemetry:
a. Raise the antenna on the PRM to its upright position.
b. For optimum ZIP telemetry communication, position the PRM antenna
within 3 m (10 ft) of the pulse generator.
c. Remove obstructions between the PRM and the pulse generator.
NOTE: Reorienting the PRM a
ntenna or repositioning the PRM may improve
ZIP telemetry performance. If ZIP telemetry performance is not satisfactory,
use the telemetry wand instead.
9